5MHz OSCILLOSCOPE O-1305 INSTRUCTION MANUAL SAFETY Symbol in This Manual This symbol indicates where applicable cautionary “or other information is to be found. Power Source This equipment operates from a power source that does not apply mare than 250V rms between the supply conductors or between either supply conductor and ground. A protective ground connection by way of the grounding, condustor in the power cord is essential for safe operation. Grounding the Product This equipment is grounded through the grounding conductor of the power cord. To avoid electrical shock, plug the power cord into a properly wired ceceptacle before connecting to the equipment input or output terminals. , (SWEEP CIRCUIT] Sweep frequency : 10Hz to 100kHz (4 ranges plus fine adjustment) Sweep linearity : 5% or less Synchronization : Synchronous sweep. internal neg~ ative synchronization and exter nal synchronization Synchronization : Internal: Wave amplitude on CRT amp] i tude over | division C4MHz to SMHz: 2 divisions ar nore) External: 2Vp-p ar more CUNTENS ITY MODULATION] Sensitivity : 25Vp-p or more CSUPPLY VOLTAGE] 100/120/220/240VAC + 10%. 50/60Hz (Max. 250VAC) CPONER CONSUMPTION] Approx. 14. SW (DIMENSIONS AND WEIGHT] Width :.130mm (130mm) Height: 190mm (213mm) Depth : 280mm (307mm) Values enclosed in parentheses indicate the dimensions in- cluding protrusions Weight: Approx. 3. 5kg Instruction manual: 1 copy Input cord (CA-46): 1 Replacement Fuse : 2 [ACCESSORIES J Use the Proper Power Cord – 2 Use only the power cord and connector specified for your product. Use the Proper Fuse To avoid fire hazard. use a fuse of the correct type Do not Operate In Explosive Atmosphores ‘o avoid explosion, do not operate this product in an explosive atmosphere. Do not Remove Cover or Panel o avoid personal injury. do not remove the cover or panel. Refer servicing to qualified personnel. Voltage Change f the pawer source is not appliod to your product, contact your dealer. To avoid clectricnl shock, do not perform the voltage change. Ay prodwal or KENWOOD CORPORATION 14-6 Dogenzaka 1-chome, Shibuyneku Tokyo 100 samy FEATURES Othe vertical axis has high sensitivity of 1OmVAlly A wide frequency response ranges over 5Mlls ONC amplifiers used for both the vertleal and horizontal axis enable external low-spood awoop below IkHz. OLow power consumption and heat generation offor stable operation OThe angle of the trace line may be controllod easily on the rear panel. OSmall in size and light in weight for portability, OProvided with an external synchronization signal input terminal. SWITCHES AND CONTROLS 1, PANELS The following describes the switches and controls on the panels. See the profile shown in Figure L. [Front Panel] @ @ @ Pilot lamp The pilot lamp if oh when the CO-1308 is ready for operation. POWER gm ON/ MM OFF switch Press the switch to turn on the power (button depressed). Press it again to turn off the power Cbuttoa released) VEAT. INPUT terminal [oput vertical signal to this terminal. PRECAUTIONS FOR HANDLING L. @ “Licno) terminal Inspect for the supply voltage before using the product, The rated,voltage of the product is indicated on its rear pane}. Note that use on improper voltage will cause troubles or damages. If you make sure that the supply voltage is proper connect the power cord plug to the wall socket. ’ The preduct has jhternal circuits that carry, high voltage. Never. detach the case Do not use the produét in the following places: +A place exposed to the direct sunlight . +A place exposed to high temperature and/or high humidity ; ; *Near equipment or device which produces strong magnetic force and/or impulso voltage Do not apply voltage that excooda.the rating to any input terminal. Between VERT. INPUT terminals @ and @ : 800Â¥ CDC*RAC peak) or GOOVp-p Between EXT. SYNC, and TOR, (NPUT terminals @ and © : SOV COC*AC poak) or 100Vp-p The trace line is subject to tilt under the influence of the earth magnotism. depending on in what direction the produgt {9 places. Adjust the angle of the traco ling in the condition where the product is used. ; Connect this terminal to the the ground. EXT. SYNC/HOR. INPUT terminal Input external synchronization signal and external horizontral signal to these terminals. Connect the GND lifie to the GND terminal @. AC-DC selector saitch Selects vertical signal coupling. If this switch is set to DC input signals are DC coupled, allowing observation ~ SPECIFICATIONS [eRT} > TSARB31 [VERTICAL AXIS] Deflection : Approx. 10mV/div. sensitivity Frequency response: DC:DC to SMHz or more (-3dB) : AC:2Hz to SMHz or more (-3dB) Input resistance &: IMQ +2%. 50pP or less capacitance Attenuator Gain control Max. input vol tage 21, 1/10, 1/100 and GND (Step-to-step errar: 5% or less) « Step-less control aver 22dB : S00V COC-+AC peak? or 600Vp~p CHORIZONTAL AXiS) Deflection : 300mV/diy. or more Sensitivity Frequency response: DC to 250kKHz (with BXT.GAIN at . the maximum) DC to approx. 30kHz (with EXT. GAIN at the 1/2 point) Input resistance &: IMQ (+20%), 40pF or less capacitance (Provided the SYNC switch is set : _ ta IND Attenuator : Step-less attenuation ta 0 CEXT. GAIN) Vax. input : SOV CDC+AC peak) or 100Vp~p vol tage : of signals with DC components. If it is set to AC input signals are AC coupled, allowing observation of AC signals only containing no BC components. Â¥. ATT A vertical attenuator may be selected from 1/1, amplifier is grounded and terminal @ opens. V. GAIN control Controls the gain of the vertical axis, Adjust the wave amplitude on the CRT properly with this control, If it may not be adjusted completely, use the Â¥. GAIN and Â¥. ATT together. POSITION control Vertical position controls which move the waveform up and down on the CRT. “dh POSITION control Horizontal position controls. which move the waveform to the right and ‘left on the CRT. SWEEP RANGE selector switch Selects .the sweep frequencies. Adjust the number of wave peaks displayed on the CRT for easy observation using the SWEEP RANGE and SWEEP VARI G2. Values such as “10~100″, “100~-1k”, etc, indicate the sweep frequencies. SWEEP VARI/EXT. GAIN control A fine control of the sweep frequency and gain control of the external horizontal signals. If the SWEEP/EXT. @ is set to the SWEEP position, the SWEEP VARI/EXT. GAIN control serves as a fine control of the sweep frequency. [f the SWEEP/EXT. 9) is set to the EXT position, it controls the gain of the signal input to terminal @ and varies the horizontal amplitude on the CRT. Note that the horizontal frequency 4 Attenuation , 1/10 and 1/100. At the GND position, input signal to the vertical – SET ae ee